Product Details:
Model Name/Number | KAL 200 |
Brand | Halstrup-Walcher |
Display Type | Alphanumeric display with 2 x 20 characters, backlighting |
Weight Kg | Approx. 4.5 kg |
Hysteresis | 0.1 % of max. value |
Temperature Coefficient Zero Point | +/- 0 % (cyclical zero-point correction) |
Calibration Temperature | 22 Deg C |
Medium | Air, all non-aggressive gases |
Operating Temperature | 10 to 40 Deg C |
Storage Temperature | 10 to 70 Deg C |
Pressure calibration device with integrated pressure generator for stationary and mobile applications.
The KAL 200 calibration device combines a pressure transmitter and pressure gauge. These can be used to test and calibrate pressure sensors and other pressure gauges for low pressures both quickly and easily. The desired pressure is indicated on the digital display and pressure is generated quickly and precisely through the use of a closed loop, thereby eliminating the need for an external pressure generator.
Special Features:
High level of accuracy and long-term stability through automatic zero-point calibration Simple menu navigation for unit conversion, setting calibration points and setting the measurement range Menu navigation in German, English, Italian, French and Spanish Generation of positive and negative differential pressures Can be calibrated Easy to operate and produce calibration curves via USB port and PC
Measurement ranges:
Margin of error:
Input / power supply (test objects):
Product Details:
Model Name/Number | KAL 84 |
Brand | Halstrup-Walcher |
Display Type | Digital |
Usage/Application | Laboratory |
Accuracy | 0.2% |
Weight Kg | 3 kg |
Power Supply | 115 - 230 VAC adaptor |
Analog Output | 0 to 1 V (RL 2 k) 2 connectors 4 mm |
Pressure calibration device with integrated pressure generator for mobile applications.
Special features:
The portable, battery-operated KAL 84 calibration system combines a pressure transmitter and pressure gauge. These can be used to test and calibrate pressure sensors and other pressure gauges for low pressures both quickly and easily. A high degree of accuracy and excellent reproducibility of results combined with a robust design give the system all the functionalities required for laboratory and service applications.
Calibration in mmHg (optional) makes the system ideal for applications in accordance with the regulations for users of medical products, e.g. blood pressure monitors and infusion/suction pumps
Measurement ranges